Maintenance
143
Spray Chamber
Check for leaks around the nebulizer and drain fitting. Inspect the spray chamber for
deposits and check the condition of the O-rings.
Peristaltic Pump
Check that the pump rollers are clean, not scored, free from deposits due to spills,
and move freely. Remove and clean the pump head if necessary. Replace the pump
rollers if necessary.
Periodically observe the drain tubing. Liquid should drain smoothly from the spray
chamber. If not, adjust the pump tension on the pump tubing.
Drains
Check the spray chamber drain fittings, tubing and connections and replace if
necessary.
General System Maintenance
•
Air filters:
clean or replace the spectrometer and generator air filters as
necessary.
•
Chiller:
Flush out the chiller every 6 months. Fill the chiller reservoir with the
coolant fluid (Part No. N0776099). The chiller reservoir has a capacity of 4.2
liters (1.1 US gallons).
Refer to Chapter 4, Installation for more information on filling the chiller.
Caution
To extend the life and aid in preventing devitrification of glassware, it is
necessary to remove all traces of alkali on glassware caused by handling it
with bare hands. To remove the alkali, wipe the glassware clean with
cotton wool dipped in alcohol.
Caution
Do not use deionized or tap water in the chiller. Do not use an
algaecide other than that specified above.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......