Troubleshooting
253
Problem
Possible Cause(s)
Corrective Action
Results
erroneously low
Standard solutions may have
been improperly made.
Rerun proven standards.
Blank solution may be
contaminated or not matrix-
matched to the standards.
Remake the blank solution. Acidify the
blank if the sample matrix is acidified.
Does the matrix composition
of the samples vary from that
of the blanks and standards?
If possible, prepare the standard and
blank in the same acid matrix used for
the samples. Or, internal
standardization may be required.
Is the concentration for the
second standard lower than
that for the first?
Increase the Read Delay to flush the
sample introduction system completely.
Alternatively, aspirate a blank between
standards.
Is background correction being
used?
Check the placement of background
correction points. Refer to the
Winlab32 Software Guide
on the
WinLab CD for details
.
Are you analyzing aluminum
at 167 nm?
Run a high nitrogen purge for one hour
before starting the analysis. Refer to the
Winlab32 Software Guide
on the
WinLab CD for details
.
Undetectable emission
signal
The sample introduction
system may not be set up
correctly, or may require
maintenance.
Refer to
Cleaning the Sample
Introduction System
on page 144
.
The shutter may be closed.
Check the shutter position set in the
software. Refer to the Winlab32
Software Guide on the WinLab CD for
details
.
Memory effects
Spray chamber may not be
being drained properly, or may
need cleaning.
Check drain fittings and drain tubing.
Clean the spray chamber if necessary.
Table 6-1.
Performance Problems: Troubleshooting Table
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......