Maintenance
162
4. Carefully slide the torch through the plasma induction plates. Make sure that the
Ground Pointer is in close proximity to the end of the torch glass but
not inside
the torch glass. If needed, gently push the Ground Pointer slightly towards the
rear of the torch box. See the following figure. The torch assembly is now
installed on the instrument.
6
Locking Lug
7-9
Channels
10
Torch Mount
Item
Description