Troubleshooting
252
Problem
Possible Cause(s)
Corrective Action
Cannot obtain
suggested BEC value
You may not be comparing the
appropriate BEC value for the
wavelength you are using.
Refer to the Winlab32 Software
Guide
on the WinLab CD
for details
.
The torch position may need
adjustment.
Refer to the torch adjustment
procedures for your instrument.
The plasma viewing position
may need to be optimized.
Refer to the Lateral (radial view mode)
and X/Y (axial view mode) torch view
procedures. See
Setting the Torch
Viewing Position
on page 121
.
RF power, nebulizer argon
flow, and/or plasma viewing
height may need adjustment.
Adjust as necessary. Also, check that
you are using the recommended
settings for your instrument.
The purge windows may
require cleaning.
Refer to the instructions for removing
and cleaning the purge windows for
your instrument
.
Results
erroneously
high
Corrections may not have been
made for the reagent blank.
Aspirate a blank and rerun standards
and samples.
Standard solutions may have
deteriorated or may have been
improperly made.
Restandardize with a proven standard.
Background emission may be
present.
Use background correction.
Table 6-1.
Performance Problems: Troubleshooting Table
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......