Error Messages
269
System Error (SYSERROR) Messages for RF Generator
In the following chart, the first column lists the error code. The second column lists
the error messages that appear in the generator diagnostics software. The third
column gives the meaning of the error messages. The fourth column gives
suggestions for corrective action.
In some cases, the error value is 0 and therefore meaningless. However, in cases
where the RF generator is operating out of range, the value of the parameter is
reported. For example, if the error code is "01", Vds out of voltage range, the value
is the actual Vds voltage.
The following messages are listed numerically by error code.
Note
For every error message reported, four pieces of information are given; the
error code, the error value, the time the error occurred and the error
message.
00
No Error
There is no error.
No corrective action is needed.
01
Vds out of
voltage
tolerance
window
The drain to source
voltage across the
power FET transistor
is out of the
Tolerance window.
Contact PerkinElmer Service.
02
Ids 1 out of
level 1
window
The drain to source
current across the
power FET transistor
1 is out of the level 1
Tolerance window.
Contact PerkinElmer Service.
03
Ids 2 out of
level 1
window
The drain to source
current across the
power FET transistor
2 is out of the level 1
Tolerance window.
Contact PerkinElmer Service.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......