Error Messages
280
69
Fiber Optic
not
detecting
plasma in
the monitor
The firmware cannot
detect the plasma
after ignition. The
fiber optic will do a
hardware shutdown.
Check that the fiber optics is in line with the
plasma induction pl ates. If the problem
persists contact PerkinElmer Service.
70
Neb back-
pressure
>60 psi
(413 kPa)
& shut off
neb
The nebulizer has
been shut off once the
nebulizer back-
pressure has fallen
below 60 psi (413
kPa).
Check the sample introduction system for
leaks and blockage and correct. Check that
all the fittings are tight and check if the
nebulizer spray pattern is correct. If the
problem persists contact PerkinElmer
Service.
71
Neb heater
sensor is
not
working-
heater off
The nebulizer
compartment heater
sensor is not working
properly and the
heater is now off.
If there is a large variance in room
temperature you might experience instrument
drift if the nebulizer compartment heater is
off. Otherwise the instrument is fully
functional.
Contact PerkinElmer Service to have the
nebulizer compartment heater repaired.
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......