System Description
81
EMO Switch
The EMO switch,
Emergency Off Switch
, is the illuminated red switch on the front of
the instrument. If it is blinking slowly the instrument is in the middle of an ignition
cycle.It blinks rapidly after the Emergency Off Switch is depressed. This is an
indication that the switch has been depressed and to remind you to push the switch
again to release it. This switch shuts off the plasma in an emergency by
disconnecting the main voltage circuitry in the RF generator. To restart, you must
release the switch by pressing the switch again. Reset the RF generator using the
software Reset button. (The spectrometer stays on.)
Figure 3-10
Location of Red Emergency Plasma Off Switch.
Software Controls
Many of the hardware settings are controlled by the software.
•
RF Power:
Power levels can be adjusted in 1-watt increments.
•
Plasma and auxiliary argon flow rates:
Flow rates can be automated during the
analysis with specific flow rates for each element if desired. Plasma argon is
adjustable in 1 L/min increments. Auxiliary argon is adjustable in 0.1 L/min
increments.
Item
Description
1
Emergency Off Switch
1
Summary of Contents for OPTIMA 8000
Page 1: ...OPTIMA 8000 Customer Hardware and Service Guide ICP OPTICAL EMISSION Return to Document Menu...
Page 2: ......
Page 3: ...Optima 8000 Customer Hardware and Service Manual...
Page 12: ...Contents 10...
Page 18: ...16 Indicates the ON position of the main power switch Indicates alternating current...
Page 24: ...22 Figure C Location of warning labels in the sample compartment 1 2...
Page 28: ...26...
Page 29: ...Safety Practices 1...
Page 49: ...Preparing Your Laboratory 2...
Page 61: ...Preparing Your Laboratory 60...
Page 62: ...System Description 3...
Page 89: ...Installation 4...
Page 136: ...Maintenance 5...
Page 241: ...Troubleshooting 6...
Page 264: ...Error Messages 7...
Page 284: ......