45
IVIS
®
Spectrum Hardware Manual
8
Troubleshooting
Measured Temperature Does Not Equal the Demand Temperature . 45
Photographic Image Is Unacceptable . . . . . . . . . . . . . . . . . . 46
Luminescent Image Is Unacceptable . . . . . . . . . . . . . . . . . . 47
No Image Is Produced . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Hardware Problems . . . . . . . . . . . . . . . . . . . . . . . . . . . 48
Accessing the Area Under the Imaging Platform . . . . . . . . . . . . 50
8.1 Measured Temperature Does Not Equal the Demand Temperature
At start up, the Living Image
®
software programs the CCD camera to maintain
the CCD at -90° C. If the camera power supply and ThermoCube chiller unit
remain on, the system maintains this temperature regardless of whether the
Living Image software is open or the computer is turned on.
■
To check the temperature of the CCD, click the Temperature square (red or
green) in the Camera Control Panel of the Living Image software
.
Figure 8.1
IVIS System control panel in the Living Image software
Click the temperature square to display
the demand and measured temperatures
of the camera and imaging chamber.
Problem
Possible Cause
Corrective Action
Measured temperature is
warmer than the demand
temperature.
A problem may exist with the
camera or the ThermoCube unit.
Contact IVIS Imaging Systems Technology
Support.
Summary of Contents for Caliper IVIS Spectrum
Page 6: ...Contents iv ...
Page 10: ...1 Welcome 4 This page intentionally blank ...
Page 22: ...3 Warnings 16 This page intentionally blank ...
Page 26: ...4 Legal Notices 20 This page intentionally left blank ...
Page 30: ...5 Specifications 24 This page intentionally blank ...
Page 40: ...6 IVIS Spectrum Components 34 This page intentionally blank ...
Page 50: ...7 Operating the IVIS Spectrum 44 This page intentionally blank ...
Page 68: ...10 Servicing Maintaining the ThermoCube Chiller Unit 62 This page intentionally blank ...
Page 70: ...A Spare Parts List 64 This page intentionally blank ...
Page 74: ...Index 68 ...