33
IVIS
®
Spectrum Hardware Manual
6.11 High Reflectance Hemisphere
You can special order a high reflectance hemisphere that is used to check for
light contamination in the imaging chamber
.
The hemisphere does not emit any photons when it is imaged in a light-tight and
contaminant-free imaging chamber. As a result, it is not visible in a luminescent
image. If a light leak or an internal phosphorescent contaminant is present in the
imaging chamber, the hemisphere reflects that light and is visible in a
luminescent image. If you suspect that there is light contamination in the
imaging chamber, see
Figure 6.10
High reflectance hemisphere
Summary of Contents for Caliper IVIS Spectrum
Page 6: ...Contents iv ...
Page 10: ...1 Welcome 4 This page intentionally blank ...
Page 22: ...3 Warnings 16 This page intentionally blank ...
Page 26: ...4 Legal Notices 20 This page intentionally left blank ...
Page 30: ...5 Specifications 24 This page intentionally blank ...
Page 40: ...6 IVIS Spectrum Components 34 This page intentionally blank ...
Page 50: ...7 Operating the IVIS Spectrum 44 This page intentionally blank ...
Page 68: ...10 Servicing Maintaining the ThermoCube Chiller Unit 62 This page intentionally blank ...
Page 70: ...A Spare Parts List 64 This page intentionally blank ...
Page 74: ...Index 68 ...