59
IVIS
®
Spectrum Hardware Manual
10.4 Sliding the Tray Out
The ThermoCube unit sits on a tray that must be pulled out to access the unit.
1. Use the ¼" hex head (Allen) wrench to loosen and remove the bolt that
secures the tray under the ThermoCube unit to the console
.
2. Move any cables or tubes out of the way so that the tray can be pulled out
freely.
3. Pull out the tray with the ThermoCube unit
.
NOTICE
The ThermoCube unit has its own power switch located at the rear of the unit.
The power switch is not visible until the unit is pulled out. Make sure that the
power switch is not accidentally turned off when you slide the tray out or back
into the console.
4. Remove the fill cap on the ThermoCube chiller and check coolant level.
If
the coolant level is half way up the neck of the tank, no additional
coolant is required
.
Figure 10.3
Loosen the tray locking bolt
Figure 10.4
Removing the fill cap