2. Important Safety Instructions
8
Water & Moisture
VOLTAGE
VOLTAGE!
Do not use this product near water (for example, near a sink or
wet room) due to risk of electric shock, electrical damage, and/or failure.
Laboratory Space &
Ventilation Requirements
The IVIS
®
Spectrum requires a minimum of 100 square feet (9 square meters)
of laboratory space for adequate ventilation. At least of six inches of space is
required behind the rear flat panel to provide unobstructed air flow.
Allocate an area at least 36 inches wide and 51 inches deep for the imaging
system console to allow the door to be fully opened and provide the
recommended air space behind the instrument.
An additional area that is 48 inches wide should be provided (preferably on the
right hand side) for a table or cart to hold the computer and provide a work
surface.
2.7 Power Considerations
Power Sources
The IVIS
®
Spectrum is configured for the voltage requirements of the
installation locality that was specified at the time of order. If the system is
moved to another area, check to make sure that the same voltage requirements
exist.
An IVIS Spectrum that operates on 120VAC requires a dedicated 20 Amp
circuit that is not shared with other loads except the computer and the optional
XGI-8 Gas Anesthesia System. Make sure that all equipment is plugged into a
properly grounded AC supply.
The computer should share the same circuit to avoid ground loops. Never use
auxiliary power outlets on other equipment to supply the IVIS Spectrum. Since
the system contains internal surge protection, it must be powered directly from
the main electrical supply wall socket. Use the surge protector provided by
Caliper for the computer and other accessories.
An IVIS Spectrum configured for 230 VAC requires a dedicated minimum 10
Amp circuit subject to the same conditions noted above. For more details on the
operating requirements, see
VOLTAGE
VOLTAGE!
The IVIS Spectrum can operate at multiple voltages (100, 120,
220, 240 VAC); however, you are
not permitted
to change the input voltage
to any of the system components. Several internal modifications are required
for voltage change. If the operating voltage must be changed, contact IVIS
Imaging Systems Technology Support.
Power Cord Protection
Power supply cords should be routed so that they are unlikely to be walked on
or pinched by items placed upon or against them. Pay close attention to
receptacles and to points of connection between cords and equipment.
Lightning & Power Line
Surges
The IVIS
®
Spectrum is supplied with an internal surge protector that is not
customer serviceable. Contact IVIS Imaging Systems Technology Support
regarding equipment servicing.
Summary of Contents for Caliper IVIS Spectrum
Page 6: ...Contents iv ...
Page 10: ...1 Welcome 4 This page intentionally blank ...
Page 22: ...3 Warnings 16 This page intentionally blank ...
Page 26: ...4 Legal Notices 20 This page intentionally left blank ...
Page 30: ...5 Specifications 24 This page intentionally blank ...
Page 40: ...6 IVIS Spectrum Components 34 This page intentionally blank ...
Page 50: ...7 Operating the IVIS Spectrum 44 This page intentionally blank ...
Page 68: ...10 Servicing Maintaining the ThermoCube Chiller Unit 62 This page intentionally blank ...
Page 70: ...A Spare Parts List 64 This page intentionally blank ...
Page 74: ...Index 68 ...