57
IVIS
®
Spectrum Hardware Manual
10
Servicing & Maintaining the ThermoCube Chiller Unit
Required Tools . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Turning Off Electric Power to the Main Console . . . . . . . . . . . . 57
Removing the Console Rear Panel . . . . . . . . . . . . . . . . . . . 58
Sliding the Tray Out . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
Filling the ThermoCube Unit . . . . . . . . . . . . . . . . . . . . . . . 60
Reassembling the System . . . . . . . . . . . . . . . . . . . . . . . . 61
Restarting the IVIS Spectrum . . . . . . . . . . . . . . . . . . . . . . 61
Over time, the coolant level in the ThermoCube thermoelectric chiller unit may
diminish and require replenishing. It is recommended that you check the coolant
level in the ThermoCube unit every six months. This chapter explains how to
add coolant to the Thermocube unit.
10.1 Required Tools
The following tools and items are required, but not provided.
•
Allen Wrenches (hex head wrench): 3/32" and ¼" measured across the
flats (AF).
•
Measuring cup with a spout.
•
Spectral TE Camera Cooling Mixture (part no. 119219, includes a 3/32"
and ¼" allen wrench) or a solution of 25% pure ethanol, 75% distilled
water.
•
Paper towel or cloth for clean up.
10.2 Turning Off Electric Power to the Main Console
VOLTAGE
VOLTAGE!
Do not remove any panels from the console until you turn off the
IVIS Imaging System disconnect the power cord. Otherwise, electric shock
could result.
1. Pull the front toggle switch out and down to the standby position.
2. Turn off the main power switch on the rear panel and disconnect the console
power cord from the wall socket.
3. Move any cables away from the lowest rear panel of the console so that the
panel can be removed
. (No other cables need to be removed.)
Summary of Contents for Caliper IVIS Spectrum
Page 6: ...Contents iv ...
Page 10: ...1 Welcome 4 This page intentionally blank ...
Page 22: ...3 Warnings 16 This page intentionally blank ...
Page 26: ...4 Legal Notices 20 This page intentionally left blank ...
Page 30: ...5 Specifications 24 This page intentionally blank ...
Page 40: ...6 IVIS Spectrum Components 34 This page intentionally blank ...
Page 50: ...7 Operating the IVIS Spectrum 44 This page intentionally blank ...
Page 68: ...10 Servicing Maintaining the ThermoCube Chiller Unit 62 This page intentionally blank ...
Page 70: ...A Spare Parts List 64 This page intentionally blank ...
Page 74: ...Index 68 ...