Pulscon LTC57 PROFIBUS PA
Diagnostics and troubleshooting
2015
-0
3
79
262
Module connection
1. Check module connection
2. Change electronic modules
F
Alarm
270
Main electronic failure
Change main electronic module
F
Alarm
271
Main electronic failure
1. Restart device
2. Change main electronic module
F
Alarm
272
Main electronic failure
1. Restart device
2. Contact service
F
Alarm
273
Main electronic failure
1. Emergency operation via display
2. Change main electronics
F
Alarm
275
I/O module failure
Change I/O module
F
Alarm
276
I/O module failure
1. Restart device
2. Change I/O module
F
Alarm
282
Data storage
1. Restart device
2. Contact service
F
Alarm
283
Memory content
1. Transfer data or reset device
2. Contact service
F
Alarm
311
Electronic failure
1. Transfer data or reset device
2. Contact service
F
Alarm
311
Electronic failure
Maintenance required!
1. Do not perform reset
2. Contact service
M
Warning
Diagnostic of configuration
410
Data transfer
1. Check connection
2. Retry data transfer
F
Alarm
412
Download process
Download active, please wait
C
Warning
435
Linearization
Check linearization table
F
Alarm
437
Configuration
incompatible
1. Restart device
2. Contact service
F
Alarm
438
Dataset
1. Check data set file
2. Check device configuration
3. Up- and download new configura-
tion
M
Warning
482
Block in OOS
Set block in AUTO mode
F
Alarm
484
Simulation failure mode Deactivate simulation
C
Alarm
485
Simulation measured
value
Deactivate simulation
C
Warning
494
Switch output
simulation
Deactivate simulation switch output
C
Warning
Diagnostic
number
Short text
Repairing action
Status signal
(from the
factory)
Diagnostic
behavior
(from the
factory)