Pulscon LTC57 PROFIBUS PA
Operating menu
15.4.7
Data logging submenu
Data logging
Assign channel 1 to 4
Note:
The
Data logging
submenu is only available if the advanced functionality of
the HistoROM has been activated in the device.
Navigation
Diagnostics
Data logging
Blocking
Navigation
Diagnostics
Data logging
Assign chan. 1 to 4
Description
Allocate a process variable to the respective data logging channel.
Selection
• Off
• Level linearized
• Distance
• Unfiltered distance
• Terminal voltage
• Electronic temperature
• Measured capacitance
a
• Absolute echo amplitude
• Relative echo amplitude
• Absolute EOP amplitude
• EOP shift
• Noise of signal
• Calculated DC value
b
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
a
Visibility depends on order options or device settings
b
Visibility depends on order options or device settings
Factory setting
Off
Additional information
A total of 500 measured values can be logged. This means:
• 500 data points if 1 logging channel is used
• 250 data points if 2 logging channels are used
• 166 data points if 3 logging channels are used
• 125 data points if 4 logging channels are used
If the maximum number of data points is reached, the oldest data points in the data
log are cyclically overwritten in such a way that the last 500, 250, 166 or 125
measured values are always in the log (ring memory principle).
Note:
The logged data are deleted if a new option is selected in this parameter.