201
5-0
3
34
Pulscon LTC57 PROFIBUS PA
Mounting
6.2
Mounting the device
6.2.1
Required mounting tools
• For mounting thread 3/4 in: Hexagonal wrench 36 mm
• For mounting thread 1-1/2 in: Hexagonal wrench 55 mm
• To shorten rod probes: Saw
• To shorten rope probes:
•
Allen key AF3 mm (for 4 mm ropes) or AF4 mm (for 6 mm ropes)
• Saw or bolt cutter
• For flanges and other process connections: appropriate mounting tools
• To turn the housing: Hexagonal wrench 8 mm
6.2.2
Shortening the probe
Figure 6.17
Shortening rod probes
Rod probes must be shortened if the distance to the container floor or outlet cone is less than
10 mm (0.4 in). The rods of a rod probe are shortened by sawing at the bottom end.
Note!
When shortening the probe: Enter the new length of probe into the Quick Start Guide which
can be found in the electronics housing behind the display module.