201
5-0
3
134
Pulscon LTC57 PROFIBUS PA
Operating menu
Switch-on value
Figure 15.10
Blocking
Navigation
Setup
Advanced setup
Switch output
Switch-on value
Prerequisite
Switch output function
(page 132) =
Limit
Description
Enter measured value for the switch-on point.
User entry
Signed floating-point number
Factory setting
0
Additional information
The switching behavior depends on the relative position of the
Switch-on value
and
Switch-off value
parameters:
Switch-on value
>
Switch-off value
• The output is closed if the measured value is larger than
Switch-on value
.
• The output is opened if the measured value is smaller than
Switch-off value
.
A
Switch-on value
B
Switch-off value
C
Output closed (conductive)
D
Output opened (non-conductive)
t
t
A
B
C
D