Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
16
1-800-343-4048
Belts
1
Check belt tension using belt tension gauge (see figure 23). Tension gauge should indicate
minimum 45 lbs.
2
Adjust tension if required by adjusting motor bracket. Replace belts if needed.
1
Remove the plastic dust cap from the pump (see figure 27).
2
Rotate the compressor pulley until the grease fitting is visible in the dust cap hole.
3
Using a grease gun extension adapter to engage the grease fitting, supply the proper amount
of grease according to the grease delivery Table 1, figure 27.
Note: Pump grease gun before feeding to eliminate air from the needle adapter. Each pump
equals 0.65 grams of grease. The volume of grease is less after the 2nd pump since some
grease remains in the extension adapter.
4
Replace the plastic dust cap.
Grease Orbit Scroll
Compressor Bearings
Cleaning
If necessary, wipe the compressor or generator with a clean, dry cloth on an as-needed
basis. Do not use water, aerosols or other cleaning agents on the units. Use of any liquid
detergent to clean the units could pose an electrical hazard.
Schedule 1 Instructions
View the video, Scroll Compressor Service Maintenance, prior to performing the following tasks.
The detail steps are shown in the video instructions.
1
Remove the top panel on the unit by removing two screws with a flat head screwdriver.
2
By hand, push-in and rotate the intake filter cap counter clockwise to remove (figure 26).
3
Replace the old filter element with a new one, reverse steps to reassemble.
Figure 26 Filter Element
Intake Filter
2