Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
15
1-800-343-4048
Figure 25 Tri-Gas Expanded View Replacement Parts
(One unit shown)
(Refer to Page 27 for list of parts): Front View
Figure 24B Tri-Gas Maintenance Items: Back View
(One unit shown)
Schedule 3 (refer to figures 24A and 24B)
Replace Quantity Hours
1
Frequency
2
Filter Element
1
5,000
1 year
(P/N 100-09-DX)
Filter Element
1
5,000
1 year
(P/N 100-09-BX)
Filter Element
4
3
5,000
1
year
(P/N 050-05-BX)
Carbon Module
2
5,000
1 year
(P/N B06-0051)
Carbon Module
3
5,000
1 year
(P/N B06-0194)
Tube
Assembly 3
5,000
1
year
(BNGS100-120-95
Tower Module
1
15,000
3 years
(P/N 76811)
Silencer
3
15,000
3
years
(P/N BN75117)
1
Frequency hours for
Model LCMS-SF5000
is recorded by an hour meter. The hour meter is
located on the compressor control panel.
Note: The meter indicates 00000.0 hours. The hour meter does not reset.
2
Frequency for
Models LCMS-SF5001T and LCMS-SF5001NT
is by calendar time (No Hour
Meter).
3
3 elements only for
Model LCMS-SF5001NT
.
4
A - Environmental Pack
B - Horizontal Compressor
C - 2-Belt Drive on Vertical Compressor
Schedule 3
Tri-Gas Generator
For Serial Numbers
LCMS-SF5000NA: 1012A and above
LCMS-SF5001NT: 1001A and above
LCMS-SF5001T: 1001A and above
(includes all S/Ns ending in letters A,
B, or C)
4
Figure 24A Tri-Gas Maintenance Items: Front View (One unit shown)