Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
14
1-800-343-4048
Figure 22 Compressor Intake Filter
Figure 23 Compressor Two Belts
Schedule 2 (refer to figure 24A)
Replace Quantity Hours
1
Frequency
2
Filter Element
1
5,000
1 year
(P/N 100-09-DX)
Filter Element
1
5,000
1 year
(P/N 100-09-BX)
Filter Element
4
3
5,000
1
year
(P/N 050-05-BX)
Carbon Module
2
5,000
1 year
(P/N B06-0051)
Tri-Carbon Scrubber
4
1
5,000
1
year
(P/N B03-0262)
Tower Module
1
15,000
3 years
(P/N 76811)
1
Frequency hours for
Model LCMS-SF5000
is recorded by an hour meter. The hour meter is
located on the compressor control panel.
Note: The meter indicates 00000.0 hours. The hour meter does not reset.
2
Frequency for
Models LCMS-SF5001T and LCMS-SF5001NT
is by calendar time (No Hour Meter).
3
3 elements only for
Model LCMS-SF5001NT
.
4
If optional scrubber is installed on outlet gas lines.
5
Schedule 2 or 3 based on generator Serial Number. S/N ending with revision A, B, or C, go to
Schedule 3. S/N ending with no letter, go to Schedule 2.
Schedule 1
Compressor
Schedule 2
Tri-Gas Generator
For Serial Numbers
LCMS-SF5000NA: 1001 to 1011
Maintenance tasks for the Tri-Gas Generator should be performed by trained personnel familiar
with the service and safety precautions of electromechanical devices to avoid injury or damage.
Parker highly recommends service be performed by a Parker trained technician only. Call
the factory about PMP, the Preventative Maintenance Program.
Prior to servicing the Tri-Gas Generator, turn off the compressor (or compressed air supply) and
power supplies. Allow the system a minimum of 2 hours to depressurize and cool down. See
shutdown section on page 13.
To ensure product performance and reliability, periodic preventative maintenance
must
be per-
formed according to the following schedules (refer to the replacement parts section for the part
numbers when ordering).
Maintenance
Model Schedule Frequency
1,2
LCMS-SF5000
1 and (2 or 3)
5
1 year or 5,000 hours
LCMS-SF5001T
2 or 3
5
1 year
LCMS-SF5001 NT
2 or 3
5
1 year
Schedule 1 (refer to figures 22 and 23)
Replace Quantity Hours
1
Intake Filter
1
1 year or 5,000 hours
Belts
2
1 year or 5,000 hours
Tip Seal Set
1
1 year or 5,000 hours
Grease
1
1 year or 5,000 hours
Pump
1
20,000 hours