Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
24
1-800-343-4048
Figure 46 Dryer Membrane
Figure 45 Inlet / Outlet
Figure 49 Disconnect Electrical at PCB
Figure 48 Disconnect Electrical
Figure 47 Nut
The Tri-Gas Generator fuses are located in the power receptacle on the top, rear side of the gen-
erator.
Before servicing the fuses, turn the generator off and disconnect the power cord
from both the power supply and the generator power receptacle.
To access the fuses, use a small screwdriver to remove the fuse holder located in the power re-
ceptacle of the generator (see figure 50). Insert screwdriver blade in holder tab and push in and
draw out. In the generator, both the phase and neutral are fused separately. As a result of this
configuration, both fuses should be checked any time fuse replacement is warranted. Replace
either one or both fuses as necessary and re-assemble.
To maintain the safety and perfor-
mance integrity of the product, use only fuses of the size and type detailed in the Specifi-
cations section of this manual.
Figure 50 Tri-Gas Fuse
Fuse Replacement