Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
25
1-800-343-4048
System Specifications
LCMS-SF5000
LCMS-SF5001T, LCMS-SF5001NT
CSA Certification Standard
CAN/CSA C22.2 No. 61010-1
same
IEC Standard
IEC 61010-1: 2001
same
CENELEC Standard
EN 61010-1: 2001
same
UL Standard
UL 61010A-1: 2002
same
IEC 61010 Installation
Category II
same
IEC 61010 Pollution
Degree 2
same
EMC Compliance
EN61326-1: 1997/CISPR 11: 1990/IEC 1000
same
Curtain Gas (N2) Purity
99.999% Organics, 99% Oxygen
See Note 2
Curtain Gas Max. Flow/Pressure
12 SLPM/80 PSIG (5.5 Barg)
same
Max. Inlet Hydrocarbon Content
100 PPM
same
Source Gas (UHP Zero Grade Air) Purity
<0.1 PPM Total Hydrocarbons
See Note 2
Source Gas Max. Flow/Pressure
40 SLPM/110 PSIg (7.6 Barg)
See Note 2
Exhaust Gas (Dry Air) Purity
-40°F (-40°C) ATM Dew Point
See Note 2
Exhaust Gas Max. Flow/Pressure
16 SLPM/60 PSIG (4.1 Barg)
same
Suspended Liquids
None
same
Particles > 0.01 Micron
None
same
Phthalates
None
same
Sound Level
< 49 dB(A)
< 30dB(A)
Min./Max. Operating Pressure
110/140 PSIG (7.6/9.6 Barg)
3
85/140 psig (5.8/6.9 barg)
2, 3
Flow Capacity
See Pressure/Flow Charts, page 28
same
Inlet Air Consumption
180 SLPM (6.3 scfm)
same
Inlet Air Temp. Range
60°F-95°F (15°C-35°C)
same
Min./Max. Ambient Temp.
60°F/90°F (16°C/32°C)
same
Max. Ambient Relative Humidity
80%
same
Altitude 2000M
same
Heat Dissipation - Compressor
10246 BTU/Hour
-------
Heat Dissipation - Tri-Gas Generator
1229 BTU/Hour
same
Electrical Requirements
1
- Compressor 60Hz
195-254VAC, 60Hz, 1 Phase, 14A
4 -------
Electrical Requirements
1
- Compressor 50Hz
230VAC, 50Hz, 1 Phase, 13A
-------
Electrical Requirements
1
- Gas Generator
100-120/200-240VAC, 60/50Hz, 1 Phase, 4A/2A -------
Power Consumption - Compressor
13A@230VAC, 3000 Watts
-------
Power Consumption - Tri-Gas Generator
3A@120VAC, 360 Watts
same
1.5A@230Vac, 360 Watts
same
Start-Up Time
45 Minutes
same
Compressor Outlet - Compressed Air
3/8” Tube
-------
Compressor Outlet - Tank Drain
1/4” Tube
-------
Tri-Gas Generator Ports - Comp Air Inlet
3/8” Tube / 3/8” BSPT Adapter
same
Tri-Gas Generator Ports - Drain
1/4” Tube
same
Tri-Gas Generator Ports - Gas Outlets (3)
1/4” SS Compression / 6 mm adapters
same
Compressor Fuses - Primary FU1, FU2
10.3 x 38.1mm, Type T, 600V, 1.25A
-------
Compressor Fuses - Secondary FU3
6.3 x 31.7mm, Type T, 250V, 1.25A
-------
Tri-Gas Fuses - 120V
5 x 22mm, Type T, 250V, 4A
same
Tri-Gas Fuses - 230V
5 x 22mm, Type T, 250V, 2A
same
Dimensions LCMS-SF5000
34”L x 61”W x 43”H
(86cm x 155cm 109cm)
-------
Dimensions LCMS-SF5001T
-------
61”L x 34”W x 43”H
(155cm x 86cm x 109cm)
Dimensions LCMS-SF5001NT
-------
21”L x 46”W x 41”H
(53cm x 116cm x 104cm)
Product Weight LCMS-SF5000
788 lbs (357 kgs)
-------
Product Weight LCMS-SF5001T
-------
448 lbs (203 kgs)
Product Weight LCMS-SF5001NT
-------
220 lbs (100 kgs)
Shipping Weight LCMS-SF5000
1166 lbs (529 kgs)
-------
Shipping Weight LCMS-SF5001T
-------
688 lbs (312 kgs)
Shipping Weight LCMS-SF5001NT
-------
450 lbs (204 kgs)
Notes:
1. Main supply voltage fluctuations not to exceed ± 10% of nominal voltage.
2. Optimum minimum inlet pressure is 110 psig (7.6 barg) to achieve published gas specifications of instrument manufacturers.
3. Consult factory for performance specifications at lower inlet pressures, or high elevation installations.
4. PCS motors installed after 11/3/09, starting with Serial Number LCMS-SF5000NA-1010.
System Specifications