Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
7
1-800-343-4048
Figure 5
Figure 6
Location of the generator with proper ventilation is critical. Install the generator in a clean, well ven-
tilated area. To provide adequate ventilation and servicing for the generator, a minimum clearance
distance of 30 inches in front and in the rear is required.
For the LCMS-SF5000
Installation, each
of the two Tri-Gas generator cabinets is located next to the compressor cabinet (see figure 7). If
the intended location is tight for space, the compressor cabinet can be located in a remote location.
Fresh intake ventilation for the compressor cabinet enters from the bottom and sides, while warm
exhaust air exits from the right, rear side. A minimum of 3 inches clearance is required on the left
side of the compressor cabinet (see figure 8).
For the LCMS-SF5000 or LCMS-SF5001
, ventilation
for the Tri-Gas generator cabinet is from the bottom. No clearance distance is required from either
the left or right sides.
For the LCMS-SF5001NT
, the intake ventilation enters from the bottom, while
the warm exhaust air exits out the rear. No clearance distance is required from either side.
Location
Figure 7 Compressor and two Tri-Gas Generators
Location
Figure 8 Compressor
Left Side Intake