![Parker Balston Super Flow Tri-Gas Generator Series Installation, Operation And Maintenance Manual Download Page 20](http://html1.mh-extra.com/html/parker-balston/super-flow-tri-gas-generator-series/super-flow-tri-gas-generator-series_installation-operation-and-maintenance-manual_3703420020.webp)
Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
20
1-800-343-4048
Coalescing Prefilters
Schedule 2 and 3 Instructions
These instructions describe the replacing of the two coalescing pre-filters, the four in-line filters,
the two carbon modules, and the zero air tower module in the Tri-Gas generator.
1
Remove the front panel from the Tri-Gas generator by rotating the latch a quarter turn counter
clockwise with a screwdriver.
2
The replacement elements are 100-09-DX (first stage) and 100-09-BX (second stage). It is
very important to install the proper grade filter into the proper housing. Take extra care to
install in the proper sequence (see figure 30).
3
Using a 1” wrench, remove the tie nut at the base of the bowl (see figure 31).
4
Twist the bowl loose from the head of the assembly and remove carefully so as not to dam-
age or lose the o-ring (see figure 32).
5
Unscrew the element retainer at the base of the element, and replace the spent element with
a new one (see figure 33).
6
Reverse the steps to re-assemble. Make sure the o-ring is sealed in the groove before re-
installing the bowl. Hand tighten.
Figure 31 Tie Nut
Figure 32 Remove Bowl
Figure 34 Filter Bowl
Figure 33 Remove Element
Figure 35 New Element