![Parker Balston Super Flow Tri-Gas Generator Series Installation, Operation And Maintenance Manual Download Page 9](http://html1.mh-extra.com/html/parker-balston/super-flow-tri-gas-generator-series/super-flow-tri-gas-generator-series_installation-operation-and-maintenance-manual_3703420009.webp)
Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
9
1-800-343-4048
Figure 9 Compressor 20A 250V
Receptacle for NA model
Figure 10 Tri-Gas Generator Voltage Selection
(one Tri-Gas Generator shown)
Each Tri-gas generator is shipped with a tag at the power receptacle which specifies the factory
setting of the voltage selector. For optimal performance of the
EU Model
generator, the cus-
tomer should set the voltage selector to match the local power supply. To change the setting on
the voltage selector, simply remove the selector/fuse drawer from the receptacle using a small
screwdriver, turn the selector so the desired voltage (220, 230 or 240) shows in the window, and
reassemble (see figure 10). To connect the generator to the power supply, simply plug the female
end of the electrical cord into the receptacle of the generator, and the opposite end into the ap-
propriate earthed grounded power receptacle.
Before plugging the power cords into the two Tri-Gas Generator power receptacles, check the
voltage selector setting. The voltage setting must match the local power supply voltage.
Note:
“NA” models of the generator are designed for 120VAC operation only.
There are eleven plastic tubing connections for the LCMS-SF5000 Model: (2) inter-connection;
(3) drain connections; and (6) outlet gas connections. The inter-connection tube requires a length
of PFA plastic 3/8” tubing connecting the outlet of the compressor cabinet to the inlet of the Tri-
Gas Generator (see figure 11). Press the tubing into the quick-connect fittings. Assure tubing is
full inserted (about 1/2”). See Recommended Accessories for installation kits on page 10. For
remote locations additional length tubing is provided.
Piping
Figure 11 Compressor and
Tri-Gas 3/8” Connection
(one Tri-Gas Generator shown)
Figure 12 Compressor Drain to
Disposal Container
(one Tri-Gas Generator shown)
1