Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
19
1-800-343-4048
Caution! Do not attempt to remove the orbit scroll from the housing.
9
After installing half of the low pressure seal,
carefully
remove the seal from the channel to
make sure the seal is properly locking onto the channel indentations located just past the high
pressure seal.
Note:
The indentations are machined into the seal channel to prevent the low pressure seal
from moving.
10
Blow off any dust caused by removing the seal from the channel.
11
Install the low pressure seal completely.
Note:
Make sure the side and bottom lip notches are facing INWARD and DOWN into the seal
channel.
Note:
The lip notches must not be distorted in the seal or torn off.
12
Install backup tube in the dust seal channel.
13
Place dust seal over the backup tube.
Caution! The backup tube must meet at the bottom of the housing in the six o’clock posi-
tion. The dust seal must meet on the right side of the housing at the three o’clock posi-
tion.
14
Install the stationary scroll housing onto the orbiting scroll housing and reassemble the unit.
15
Tighten the pulley bolt (M8) to 265 in -lbs, using a torque wrench.
16
Tighten the six housing nuts to 265 in-lbs, using torque wrench.
There are three fuses located in the compressor cabinet (see figure 29). Before servicing the
fuses, turn the compressor off and disconnect the power cord from the power supply.
To access the fuses, remove the service panel and locate the fuse blocks. Both lines 1 and 2 are
fused separately. Remove both line fuses with a small screwdriver. Replace either one or both
fuses as required.
To maintain safety, use only fuses of the size and type specified in the
Specifications section of this manual.
Figure 29 Location of Fuses
Figure 30 BX Filter on the Left, DX Filter on the Right
Fuse Replacement
Compressor
3