Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
22
1-800-343-4048
3
Remove black velcro strap from one carbon module (see figure 40).
4
Remove tubing from top of module (see figure 41). Remove tubing from bottom of module.
5
Pull out module and replace new. Insert tubing, top and bottom.
6
Perform steps 3 through 5 for the other two carbon modules.
1
With both hands, rotate the filter bowl counter clockwise (looking from the bottom), remove
and set aside (see figure 42).
2
Rotate the filter element retainer counter clockwise (see figure 43), remove and set aside.
3
Pull out element and replace with new. Snug up retainer.
4
Re-assemble the bowl and tighten to hand-tight.
5
Perform steps 1 through 3 for the other two outlet filters.
6
Before putting on the back panel cover (see figure 44), use a soap or a leak detecting solution
to check for leaks when re-starting the generator.
Figure 38
Figure 39
Figure 40
Figure 41
Outlet Filters