Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
28
1-800-343-4048
Nominal conditions for the charts are Temperature 68°F (20°C) and Ambient Pressure 14.7 psi [1013 mbar(a)].
The maximum pressure specification for the
DRY AIR
is 60 psig. Flow meter ac/- 0.2 (Dewpoint -40°F).
Shown flow rate is for one Tri-gas generator. Double the flow rate for the total system output.
The number in the charts is from the ball on the flow meter scale. The number has no uint of measure.
Flow Rate
Pressure (PSIG)
(LPM)
20
30
40
50
60
2
1.20 1.10 1.00 .85 .70
3
1.75 1.40 1.40 1.20 1.02
4
2.30
1.90
1.85
1.60
1.42
5
2.90
2.60
2.36
2.05
1.84
6
3.45
2.90
2.85
2.50
2.27
7
4.00
3.40
3.30
2.90
2.66
8
3.90 3.80 3.37 3.10
9
3.80
3.50
10
3.90
The maximum pressure specification for the
NITROGEN
is 80 psig. Flow meter ac/-0.5 (99% purity).
Shown flow rate is for one Tri-gas generator. Double the flow rate for the total system output.
Flow Rate
Pressure (PSIG)
(LPM)
20
30
40
50
60
70
80
2
1.28 ------- ------- ------- ------- ------- -------
3
1.75 1.75 1.40 ------- ------- ------- -------
4
2.30 2.25 1.85 1.75 1.42 ------- -------
5
2.90
2.75
2.36
2.25
1.84
2.00
1.42
6
3.45
3.25
2.85
2.75
2.27
2.25
1.80
7
4.00
3.75
3.30
3.25
2.66
2.75
2.16
8
5.00
4.25
3.80
3.60
3.10
3.25
2.55
9
5.75 4.75 4.25 4.00 3.50 3.50 2.92
10
6.50
5.50
4.75
4.25
3.90
3.75
3.30
11
7.00
6.25
5.25
5.00
4.75
4.25
3.70
12
7.75
6.75
6.00
5.50
5.00
4.75
4.50
13
8.50
7.50
6.50
6.00
5.75
5.00
5.00
14
9.00
8.00
7.00
6.50
6.00
5.50
5.25
15
9.75
8.50
7.75
7.00
6.50
6.00
5.75
16
9.25
8.00
7.50
7.00
6.50
6.25
17
9.75
8.50
8.00
7.50
6.75
6.50
18
8.50 8.50 8.00 7.25 7.00
19
9.50 9.25 8.50 7.75 7.50
20
9.75 9.00 8.50 7.75
21
9.50
8.25
8.00
22
10.00
8.75
8.25
23
9.00
8.75
24
9.25
9.00
25
9.75
9.25
26
10.00
9.75