Bulletin TI-LCMS-SF5000J
Installation, Operation and Maintenance Manual
Super Flow Tri-Gas Generators
www.labgasgenerators.com
12
1-800-343-4048
Each cabinet has its own automatic water drain system. The compressor drain operates every 40 min-
utes, opens for 2 seconds. In the Tri-gas generator, the two filter bowls drain every 5 minutes and the
receiver tank drains every 30 minutes for less than one second.
The LCMS-SF5000 has built-in system diagnostics to monitor the operation of the generator and to
alert the operator both visually and audibly in case service is required.
The compressor cabinet has two LED indicators (see figure 17): Power ON, and Temperature High.
The green LED remains illuminated during normal operation. The other LED indicator, “Temperature
High” remains unlit if the compressor is functioning properly. Red illumination of the Temperature High
LED signals a faulty condition and warns the operator that the unit requires servicing.
The compressor has an hour meter to record the
accumulated operating time
of the compressor for
preventative maintenance scheduling of the serviceable components. The meter indicates 00000.0
hours. The hour meter is not re-setable.
The compressor output pressure gauge indicates the range of system pressure during operation, nor-
mally between 110 psig and 140 psig.
See the troubleshooting section of this manual on pages 30-32 to resolve any abnormal or faulty condi-
tions.
Each Tri-gas generator cabinet is equipped with three LED indicators: Low Pressure, System Ready,
and Low Temperature (see figure 18). There is also an audible alarm located on the PCB assembly
inside the cabinet.
Diagnostics
Compressor
Figure 18 Gas Generator LED Indicators
Figure 17 Compressor Cabinet LED Indicators
Tri-Gas Generators