Chapter 4. Programming Advanced Features
22
Aurora H2O User’s Manual
When a User Calibration is active, an asterisk (*) will appear in the upper left corner of
the primary display.
Before the calibration can be performed, the
Aurora H2O
must be operating correctly
in methane or natural gas, with methane selected as the
Background Gas
.
Choose
Settings...
from the
Main Menu
. A
User Cal...
selection is now provided.
Note:
If the
Aurora H2O
is set for Nitrogen as the background gas, the
User Cal
selection will be disabled/grayed out.
The
Aurora H2O
will prompt for the
User Passcode
[2719]
By default, the
Aurora H2O
is operating with the
Factory
calibration, so the
User Calibration
Menu
is disabled with the exception of the
Status
selection.
Select
Status
, and choose
User
as the
Active Calibration
.
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......