Aurora H2O User’s Manual
11
Chapter 6. Maintenance
6.2.2 Instructions for Optical Absorption Cell Style 2 (cont.)
13.
Using tweezers, insert the mirror into the hole on top of the shim, with the gold-plated side facing up. Thread the
end cap into the bottom of the absorption cell tube. Secure the threaded end cap in place with a 5/8” socket
wrench (see
Figure 97
below).
Figure 97: Replacing Mirror End Cap Assembly into the Absorption Cell
14.
Re-establish power to the
Aurora H2O
analyzer.
15.
Re-establish the flow through the sample system.
16.
If the
Aurora H2O
still indicates a
Check Mirror
fault after normal start-up, contact BH Technical Support for further
assistance.
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......