Aurora H2O User’s Manual
11
Chapter 2. Installation
2.7.2.2
Installing the Insertion Membrane Liquid Separator/Pressure Regulator Assembly (cont.)
5.
Rotate the probe counterclockwise until the pins are to the far right in the horizontal slot. At this point, the probe is
sealed against the housing interior wall. The pins will be in the middle of the second vertical slot (see
Figure 20
below).
Figure 20: Rotating the Probe
6.
Loosen the insertion nut until it is above the top of the second vertical slot (see
Figure 21
below). The probe should
not rise to the top of the second vertical slot. If the probe rises in the slot, the foot valve o-ring may have been
damaged or attacked by the process.
Note:
The threaded nut on the housing ensures that if all other safety procedures are disregarded, it is
mechanically impossible to remove the probe. Perform the next step regardless of the status of the foot
valve o-ring.
Figure 21: Loosening the Insertion Nut
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......