Aurora H2O User’s Manual
15
Chapter 2. Installation
2.8 Making Electrical Connections (cont.)
Note:
Supply connection wiring shall be rated at least 10°C above the rate maximum service temperature of 85°C,
be stripped back 5/16 in. (8 mm) and torqued to a minimum of 4.4 in. lb. (0.5 Nm).
3.
Run the AC power connections to the Power Terminal Block shown in
Figure 26
below. It is recommended to use
12-18 AWG (3.3 - 0.82 mm
2
) power wiring.
Figure 26: Power Terminal Block
4.
Use wiring conduit runs, separate from the
Aurora H2O
main power, for all I/O (Input/Output) leads. Wire up to
three 4-20mA outputs to the terminals labeled A, B, and C. The three analog outputs A, B and C (0-20mA or
4-20mA) are internally powered by the
Aurora H2O
. Use shielded 18-22 AWG (0.82–0.33 mm
2
) twisted pair wire,
and ground the shield at one end only. Wire up digital communications to Port 1 and/or Port 2 as labeled.
5.
Either digital port may be configured for RS-232 or RS-485. Port 1 is designated as “SCADA.” Port 2 is designated as
“SERVICE.” For operation on RS-485, 2-wire, half-duplex bus, attach the RS-485(+) to (+), and the RS-485(–) to (–
). Per Modbus-IDA specifications, a third, common conductor must also interconnect all the devices of the bus.
Attach the common conductor to RTN.
Note:
For an RS-485 Multi-drop Network, a terminating resistor must be installed across the
Aurora H2O
RS-485
terminals, or an internal terminating resistor can be applied. See below.
• When using the
Aurora H2O
in RS-485 mode, and to prevent signal reflections on the high-speed RS-485
connections, it is recommended that the far end of the RS-485 lines be terminated properly. The termination
can be accomplished in one of two ways:
•
Connect 120
, 1/4 W leaded resistors across the + and – terminals of ports 1 and 2 (both ports or
whichever one will be in use)
•
Using long-nose pliers, move jumpers J15 and J16 from pins 2 and 3 (default setting from factory) to pins
1 and 2 (see Figure 27 on page 16). J16 is the termination for port 1 and J15 is the termination for port 2. It
is also recommended that basic ESD precautions such as grounded wrist straps be used for this procedure.
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......