Aurora H2O User’s Manual
7
Chapter 6. Maintenance
6.2.1
Instructions for Optical Absorption Cell Style 1 (cont.)
11.
As you line up the mirror with the measurement cell, note the “key” for properly connecting the two pieces. The
mirror will align for mounting in only one position. Rotate the mirror assembly to line up the alignment pin with the
alignment pin slot (see
Figure 91
below).
Figure 91: Reinstalling the Mirror End Cap Assembly
12.
Loosely reinstall the six retaining bolts by hand.
13.
Use an Allen wrench to tighten the six retaining bolts in an alternating or star pattern to achieve a uniform
metal-to-metal contact. Tighten the bolts just until they are snug -
DO NOT
over-tighten these bolts.
14.
Re-establish power to the
Aurora H2O
analyzer.
15.
Re-establish the flow through the sample system.
16.
If the
Aurora H2O
still indicates a
Check Mirror
fault after normal start-up, contact BH Technical Support for further
assistance.
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......