Chapter 6. Maintenance
4
Aurora H2O User’s Manual
6.2.1
Instructions for Optical Absorption Cell Style 1 (cont.)
2.
Make a visual inspection of the mirror surface (see
Figure 86
below). Record any observation of gross
contamination. If the mirror appears to be clean, do not clean it. Instead, reinstall it carefully into the system.
Figure 86: Examining the Mirror
3.
If the mirror is contaminated, the first step in cleaning it is to use the air blower to blow away any particulate
matter on the mirror surface. Place the mirror assembly on a flat surface and repeatedly puff the air blower to
force dry clean air onto the mirror surface (see
Figure 87
below).
Figure 87: Blowing Air onto the Mirror
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......