Chapter 2. Installation
10
Aurora H2O User’s Manual
2.7.2.2
Installing the Insertion Membrane Liquid Separator/Pressure Regulator Assembly (cont.)
3.
Refer to
Figure 18
below and position the membrane end of the probe above the installed housing. Slowly lower
the probe into the housing. Avoid membrane contact with the upper section of the housing. DO NOT APPLY
DOWNWARD FORCE. The probe should slide easily into the housing. Lower the probe only enough to thread the
insertion nut one complete turn of thread engagement.
Figure 18: Installing the Probe
4.
Thread the insertion nut down by hand, lowering the probe until the insertion washer pins slide to the bottom of
the first vertical slot (see
Figure 19
below).
Note:
The threaded nut on the housing ensures that if all other safety procedures are disregarded, it is
mechanically impossible to remove the probe.
Figure 19: Threading the Insertion Nut
Membrane
Housing
Slot
Membrane
Spring
Guard
Insertion
Nut
Insertion
Washer
Housing
Vertical
First
Second
Vertical
Slot
Insertion
Nut
Insertion
Washer
Pin
Housing
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......