Chapter 2. Installation
16
Aurora H2O User’s Manual
2.8 Making Electrical Connections (cont.)
Multi-drop RS-485:
For multiple
Aurora H2O
units connected in daisy-chain fashion to the RS485 interface, it is important that the
farthest unit away from the transmitting device be the only unit incorporating any termination (see
Figure 27
below).
All other units must have jumpers J15 and J16 in positions 2 & 3 (default setting from the factory). For more details on
RS-485 wiring or operation, refer to TIA/EIA-485-A Specification.
Figure 27: RS-485 Mode Termination - Install Jumpers at J15 and J16, Pins 1 and 2
Note:
The internal terminations MUST be in the default (unterminated) position for operation in RS-232 mode.
• For operation on RS-232, connect RS-232(TXD) to (+), RS-232(RXD) to (–), and RS-232(GND) to RTN (see
Figure 28
below).
Figure 28: Input/Output Connections
J16
1
2
3
J15
1
2
3
Summary of Contents for Aurora H2O
Page 1: ...panametrics com 910 317 B August 2022 Aurora H2O User s Manual...
Page 2: ......
Page 4: ...ii no content intended for this page...
Page 12: ...Chapter x Aurora H2O User s Manual no content intended for this page...
Page 24: ...Chapter 1 Features and Capabilities 12 Aurora H2O User s Manual...
Page 44: ...Chapter 2 Installation 20 Aurora H2O User s Manual...
Page 47: ...Aurora H2O User s Manual 28...
Page 65: ...Chapter 3 Operation and General Programming 18 Aurora H2O User s Manual...
Page 96: ...Aurora H2O User s Manual 36...
Page 120: ...Chapter 5 AuroraView Interface Software 24 Aurora H2O User s Manual...
Page 140: ...Chapter 7 Troubleshooting 6 Aurora H2O User s Manual...
Page 148: ...Appendix A MODBUS RTU TCP Communications 14 Aurora H2O User s Manual...
Page 166: ...4 Aurora H2O User s Manual...
Page 168: ...Warranty 6 Aurora H2O User s Manual no content intended for this page...
Page 169: ......