CHAPTER SEVENTEEN
Test set-up
164
17.5.1
DP Growth test config
The sequence of stimulus levels delivered during a DP Growth test is
important and will greatly affect the growth function obtained.
The sequence of L2 stimulus levels used in DP Growth tests are pre-set.
The sequence of L1 levels are set by the parameters
Max L2=70 L1
and
Step L2 = -5 L1
.
Field
Description
Max L2=70 L1
The highest intensity of L2 is fixed at 70dBSPL.
The highest intensity of L1 is presented here.
Range: 75-50dBSPL
Step L2 = -5 L1
The intensity level of L2 is reduced from the 70dB in
steps of 5dB. The rate at which L1 is reduced can be
the same or less than that for L2. The rate at which L1
is reduced is set here.
Range: L1 steps: 0 to -5dB (increments of 1dB)
OTOPORT
ADVANCE
ONLY