189
Mode name
Screening
High
Low
Factory
(locked)
DP Test Config
Stim
L1
65 70 65 70
Stim
L2
55 70 55 70
Noise
Reject
52 52 52 52
Noise Mode
Rapid
Standard
Standard
Standard
Time/Point
0.64s
0.64s
0.64s
0.64s
Freq Range
1.5-8k
1.5-8k
1.5-8k
1.5-8k
Max Time
90s
180s
180s
60s
Norms
High Low High Low High Low High Low
1K - 8K
OFF OFF OFF OFF OFF OFF OFF OFF
DP Stop criteria
Min DP Loops
2
2
2
2
Min
Nlo
1.2s 1.2s
1.2s 1.2s
Min
∑
OAE
Sig
OFF
OFF
OFF
OFF
Min Band Sig
-5dB SPL
-5dB SPL
-5dB SPL
-5dB SPL
Pass
Bands
3 3 3 3
Band
Settings
SNR Rqrd SNR Rqrd SNR Rqrd SNR Rqrd
1K
6 NO
6 NO
6 NO
6 NO
1.5K
6 NO
6 NO
6 NO
6 NO
2K
6 NO
6 NO
6 NO
6 NO
3K
6 NO
6 NO
6 NO
6 NO
4K
6 NO
6 NO
6 NO
6 NO
6K
6 NO
6 NO
6 NO
6 NO
8K
6 NO
6 NO
6 NO
6 NO
DP Automation
Screening
High
Low
Factory
Autostart
ON
OFF
OFF
OFF
Autostop
ON
OFF
OFF
OFF
SNR Smart
ON
OFF
OFF
OFF
NSE Smart
-12
OFF
OFF
OFF
DP Other settings
Mandatory Save
OFF
OFF
OFF
ON
DPOAE Test Setup Parameters
OTOPORT
ADVANCE
ONLY
Test set-up
Otoport OAE+ABR
User Manual for Advance and Screener
Summary of Contents for Otoport Advance