221
If the result
Cavity OK
is displayed when the test stops, the test has
passed. Save the test and perform the other QA tests if required or exit the
QA Tests Menu
screen.
On selecting
Save
,
Patient Details
are automatically entered and the test
is allocated a unique date/time stamp.
Patient Details Field
Cavity Test Default
ID
QA1
Name
QA
First
Cavity
Each test can be reviewed individually in the
QA Occlusion Test History
area.
Note:
If an artefact is reported in the test cavity, ensure that five
successful cavity tests are performed on the Otoport before
returning it to use. Refit the probe in the cavity between each
test.
QA
area
Otoport OAE+ABR
User Manual for Advance and Screener