The following table describes each field in detail:
Field Description
Units
OAE
The total OAE Signal level
dB SPL
Noise
The total Noise level
dB SPL
NLo
The amount of data accepted due to noise
n/a
being below the noise reject level
NHi
The amount of data rejected due to
n/a
noise being above the noise reject level
Stab
Stimulus stability shows the change in probe
%
fit during a test; it is calculated by comparing
the most recent stimulus with the stimulus
recorded just before data collection began
Repro
The correlation of the two OAE waveforms
%
Time
Test time
seconds
Stim
The Checkfit Stimulus level
dB pe
Probe
The probe serial number
n/a
File
The unique test file name (populated on Save)
9.3.6
Test stop reasons
When Stop criteria have been met, the test will stop automatically and a
single beep will sound for a positive OAE result or a double beep for a
negative OAE result.
When the test stops a result is displayed on the screen (example below
left). Select
OK
to accept the test stop reason. The result is then
displayed at the top of the test screen (example below right).
The test can also be terminated manually with the
Cancel
or
End
options,
as described before.
TEOAE test
Otoport OAE+ABR
User Manual for Advance and Screener