167
17.7
Stop criteria
The test stop logic is controlled in this section. Up and down arrows on the
screen indicate that other fields are available, but not currently visible. Use
the up/down arrow keys to scroll up and down the settings and highlight
a parameter to edit. The parameter variable will flash. Use the left/right
arrows to change each setting.
17.7.1
OAE Stop criteria
When
Setup Bands
is selected a pop up table enables the edit of each
band criteria. Use the up/down arrows to select the parameter to edit and
to move between the
SNR
and
RQRD
(required) columns and left/right
arrows to edit the settings. Move down at the bottom of the
SNR
column to
edit the
RQRD
column. Select
Save
to save changes to the table settings.
Select
Cancel
to discard changes (DP tests only).
See the following tables for details of the settable Stop criteria.
Test set-up
Otoport OAE+ABR
User Manual for Advance and Screener