103
Test Result
Description
DP SNR OK
The data collected has met the criteria set. The
optimum test setting will depend on your application
e.g. screening or clinical measurement
(see chapter
17
Test set-up
)
Note: The following results will only occur if a DP SNR OK is not obtained,
providing test information feedback to the user
No Valid OAE
The data collected has not met the set Stop criteria
and the test conditions were acceptable
Noisy
The noisy data collected is three times greater than
the low noise data collected
Poor Probe Fit
The final test stimulus level is outside the stimulus ok
range or the final stimulus STMOK value is < 85%
Too Few Bands
Insufficient DPs met their stop criteria
Stopped Too
The user has ended the test manually before the
Soon
required minimum amount of data has been collected
Test Stopped
A DP Zoom or DP Growth test has stopped
automatically
10.5
DPOAE
test results
The following table lists all possible test results and associated graphics.
The table gives an explanation of the circumstances under which each
result would be shown.
DPOAE test
Otoport OAE+ABR
User Manual for Advance and Screener