CHAPTER TWENTY-THREE
Care of the Otoport
238
any sign of debris that may have entered them. Replace any part of the
probe ear piece as necessary. (See chapter
22
Probe
care
for details)
Ensure your hands are cleaned thoroughly between each patient tested.
Clean the Otoport each day before a testing session, or according to local
requirements. Ensure the Otoport is cleaned if it becomes contaminated.
Clean surfaces of the Otoport with an alcohol based sterile wipe or cloth
and antiseptic fluid. Dry the Otoport with tissue immediately afterwards. Do
not allow liquid to enter the instrument and do not immerse in fluid. Do not
allow liquid to come into contact with the connection sockets. Do no poke
any materials into either the probe or charger/pc cable sockets.
If additional hygienic protection is required, use the Otoport in an infection
control sleeve. This can also be cleaned with a sterile wipe or cloth with
antiseptic fluid. The sleeves are disposable, so should be replaced weekly
or after approximately every 50 tests.
Care of the Otoport