5 SAMPLE HANDLING AND PATIENT TESTING
7. The system will now check the integrity of the
cassette and then calibrate (Fig. 5-9). For more
information about calibration, please refer to
Chapter 4 “Calibration and Quality Control”.
NOTE: The OPTI CCA-TS2 will hold calibration
for 10 minutes for all cassette types except
B-Lac. B-Lac cassettes will hold calibration
for 2 minutes. The OPTI CCA-TS2 will beep,
warning you when only 1 minute remains
on the calibration. After this time elapses,
a message will be displayed to discard the
cassette.
NOTE: If tHb/SO
2
has been disabled (see Section
3.2.2.2),
you may attach the sample at any
time during calibration and press
.
The sample will then be automatically
aspirated after calibration and the
measurement will begin.
8. You can enter patient information while
calibration is in progress by pressing
<Patient
Info>
(Fig. 5-10).
•
Press the
<Last Patient Info>
button
(Fig. 5-10) to use the last patient info as the
default for the current patient information,
if this option is enabled in
<Setup>
(see
Section 3.2.2.2).
•
This option will populate all patient info
fields with the last patient data.
If user ID security is enabled, the user ID
field will contain the user ID of the user
currently logged in. All patient information
used as the default can be edited.
•
Verify that patient ID and all other input
parameters are correct for every patient
sample measurement.
•
Press the
<Patient Info>
button to enter
new patient information or to not use the last
patient info as the default.
Fig. 5-9 Cassette Calibration
Fig. 5-10 Patient Info