2-8
Operator’s Manual – OPTI CCA-TS2
2 SETUP
•
Close the sample chamber cover (Fig. 2-20).
•
After the cover has been closed, the
instrument will automatically detect the
presence of the calibrator cassette and begin
calibration (Fig. 2-21).
•
After the calibration is complete, you will be
prompted to open the sample chamber cover
and remove the cassette (Fig. 2-22).
•
Place the calibrator cassette back into its
pouch immediately after removal from the
instrument.
NOTE: Make sure to keep the calibrator
cassette with the instrument at all times.
Fig. 2-22 Remove Calibrator
Fig. 2-21 tHb Calibration
Fig. 2-20 Close Cover