Operator’s Manual – OPTI CCA-TS2
4-7
4 CALIBRATION AND QUALITY CONTROL
9. Press the
<Level 2>
and
<Level 3>
tabs to
display the results for the respective levels.
NOTE: Results obtained are applicable to the sensor
cassette type being used for patient sample.
•
The unit automatically checks the results
against the ranges and stores the results in its
internal database.
•
For parameters within range,
<Pass>
will
be displayed and printed.
•
For parameters out of range, or if an internal
drift is detected,
<Fail>
will be displayed.
NOTE: The printout will start automatically when
the first results are displayed. This feature
may be turned off in setup (See Section
3.2.4.1). Additional information on printing
reports can be found in Chapter 6, Data
Management.
10. Press
(Fig. 4-10) to exit the results screen.
11. You will then be prompted to open the sample
chamber cover and remove the SRC (Fig. 4-11).
12. Place the SRC back into its pouch immediately
after removal from the instrument.
13. Close the sample chamber cover.
•
If the SRC test failed, gently clean the SRC,
the optics window, and the inside cover of
the SMC with alcohol and a lint-free cloth
and repeat this process. If it fails again,
refer to the troubleshooting section in
Chapter 8 of this manual.
NOTE: For application of QC Lockout, please refer
to section 3.2.1.2.
NOTE: Verify with your particular regulatory
agency and your internal policy regarding
number of levels and frequency of SRCs to
be run.
Fig. 4-10 SRC Results (2)
Fig. 4-11 Remove the cassette