4-10
Operator’s Manual – OPTI CCA-TS2
4 CALIBRATION AND QUALITY CONTROL
Fig. 4-17 Insert Cassette
Fig. 4-16 Open Cover
5.
Open the sample chamber cover by pressing
down on the red latch (Fig. 4-16)
.
6.
Tear open the cassette pouch and remove the
cassette. Wipe any excess moisture from the
cassette with a clean dry cloth.
NOTE: If the QC sample is to be introduced with a
capillary tube, remove the syringe adapter
before placing the cassette into the chamber.
7. Insert the cassette into the chamber.
Press down to ensure that the cassette is seated
properly (Fig. 4-17).
NOTE: Run cassettes immediately after opening
pouch. Do not run, if cassette has been out
of pouch for more than 2 minutes.
8. Close the SMC cover (Fig. 4-18).
Fig. 4-18 Close Cover