Operator’s Manual – OPTI CCA-TS2
3-39
3 CUSTOMIZATION
4.
Go back to the
<System>Setup>Ethernet>
screen (Fig. 3-88) and fill in the following fields
as needed:
<IP Type>
– choose static or dynamic IP Type.
<IP Address>
– static IP address of the
instrument. Enter the IP address, if IP type is
set to static. This address can be found in the IP
address field in the
<Ethernet Test>
screen
(Fig. 3-87 above).
<Subnet Mask>
– Enter the subnet mask,
if IP type is set to static. The subnet mask is
specific to your network. Ask your network
administrator if you are unsure of the value.
<Gateway>
– Enter the gateway address, if
IP type is set to static. The gateway address
is network specific. Ask your network
administrator if you are unsure of the value.
<Discovery Port>
– is used for instrument
discovery. Use for static and dynamic IP types.
The discovery port is set to 6703 by default, but
can be changed if needed.
NOTE: If static IP type is selected, the instrument
IP address and Gateway need to be on the
same network to communicate.
5. Press
to accept the changes.
6. Press
to return to the
<Setup>
screen or
to return to the main menu.
When network connection is enabled, the status bar
displays the network icon
(Fig. 3-88).
Fig. 3-88 Ethernet Settings