Operator’s Manual – OPTI CCA-TS2
3-29
3 CUSTOMIZATION
Fig. 3-67 Printer Configuration
The OPTI CCA-TS2 can be connected to an external
printer. In the
<Configuration>
menu (Fig. 3-67),
you can select the printer configuration.
1. Select the options to be enabled.
2. Press
to accept the changes.
3. Press
to return to the
<Setup>
screen or
to return to the main menu.
3.2.4.2 Defining Units
This menu lets you change the units of measure
for pressure, temperature, output resolution, total
hemoglobin, Ca
++
, Glu, BUN (urea) and Lac.
1. In the main menu, select
<System Manager>
and
<Setup>
.
2. In the
<System Setup>
menu, press the
<System>
tab, then press
<Units>
(Fig. 3-68).
3. In the
<System>
screen, select the units for the
displayed parameters (Fig. 3-69).
4. The selection for
<Resolution>
on this menu
determines the number of digits displayed and
printed past the decimal point, for all measured
parameters.
NOTE: The selection applies to patient sample
results only. By default, the resolution for
patient samples is high.
Resolution is always high for Control and
SRC results.
Fig. 3-69 System
Fig. 3-68 Select Units