Operator’s Manual – OPTI CCA-TS2
2-9
2 SETUP
Fig. 2-23 HbCal Report
OPTI CCA-TS2
HbCal Report
DD-MMM-YY HH:MM
S/N: XXXX
Version: X.XX.XXXX
User ID:
User 123
HbCal LOT: XXXXXX
Exp. Date: MMM YYYY
HbCal Date: DD-MMM-YY
Calibration Results:
Meas’d Cal’d
tHb 12.9 13.0
S02(%) 74.6 74.9
Calibration Factors:
OLD NEW
F1 1.023 1.014
F2 1.087 1.080
F3 1.089 1.094
F4 0.000 0.000
F5 0.000 0.000
G1 1.062
G2 1.087
G3 1.082
•
The unit will now begin printing the Hb
Calibration Report showing calibration
results and calibration factors (Fig. 2-23).
•
Once the Hb Calibration is complete, the
<Ready>
display will appear (Fig. 2-24)
and the analyzer is ready for operation.
•
OPTI Medical recommends that you run
controls prior to running patient samples on
a new analyzer. You must set up your OPTI
Check control lot information in your new
analyzer prior to running them. SRCs do not
require setup and can be found in the storage
compartment in the back of your analyzer.
•
Refer to section 3.2.1 of this manual for the
QC Setup procedure. Refer to section 4.5 of
this manual for QC recommendations and
instructions for running QC measurements.
Fig. 2-24 Ready Screen